Only medical facilities

Duotrol® Normal – Quality Control Serum for Routine Testing

EAN: 04260481081868
Article number
107401
Packaging unit
1 x 5 mL Lyophilisat

Lyophilized quality control serum for routine clinical chemistry testing at the lower control level.

€12.34
Excl. 19% VAT , excl. Shipping Cost

Duotrol® Normal – Quality Control Serum for Routine Clinical Chemistry Testing

Duotrol® Normal is a lyophilized quality control serum designed to monitor the precision and trueness of quantitative measurements in clinical chemistry. It supports medical laboratories in their daily internal quality control procedures and helps maintain consistent analytical performance across a broad range of routine tests.

As the lower of two complementary control levels, Duotrol® Normal provides a defined concentration range for monitoring clinical chemistry assays. The material covers numerous analytes, including enzymes, electrolytes, metabolites, lipids and proteins.

Based on defibrinated human plasma, Duotrol® Normal is intended for use with clinical chemistry analyzers. Its manufacturer-independent design makes it suitable for different laboratory environments, provided that the relevant method-specific target values and control ranges are applied.

Why Is Routine Quality Control Essential in Clinical Chemistry?

Clinical chemistry results play an important role in diagnosing diseases, monitoring treatment and assessing changes in a patient's condition. Reliable laboratory results depend on analytical methods that consistently perform within established quality requirements.

Internal quality control allows laboratories to assess two essential characteristics of analytical performance:

  • Precision: The agreement between repeated measurements performed under specified conditions.
  • Trueness: The closeness of the average measurement result to an assigned reference or target value.

Regular control measurements help laboratories identify unexpected changes in assay performance. Such changes may arise following reagent lot replacement, recalibration, instrument maintenance or other modifications to the analytical process.

Duotrol® Normal provides a consistent control material for monitoring these factors within its assigned concentration range.

Monitoring Analytical Performance in Daily Laboratory Workflows

Duotrol® Normal is designed for routine use in laboratories performing quantitative clinical chemistry tests.

The reconstituted control serum is analyzed using the laboratory's established procedures. The results are compared with the applicable target values and control limits to determine whether the analytical system is performing as expected.

This process supports the identification of random variation, systematic shifts and trends that may affect measurement reliability.

Routine monitoring is particularly important in laboratories using automated analyzers to measure numerous parameters throughout the working day. A multiparameter control serum allows several assays to be assessed using a single control material.

Which Clinical Chemistry Parameters Can Be Monitored?

Duotrol® Normal covers a broad selection of analytes commonly measured in medical laboratories. The available parameters and assigned values depend on the analytical method and instrument.

Liver Enzymes and Related Biochemical Markers

Enzyme measurements are widely used to assess hepatocellular injury, cholestasis and other clinical conditions.

Depending on the analytical system, Duotrol® Normal provides control values for parameters such as:

  • Alanine aminotransferase (ALT)
  • Aspartate aminotransferase (AST)
  • Gamma-glutamyl transferase (GGT)
  • Alkaline phosphatase (ALP)
  • Lactate dehydrogenase (LDH)
  • Glutamate dehydrogenase (GLDH)
  • Cholinesterase
  • Bilirubin

Monitoring these assays helps laboratories assess their analytical stability during routine testing.

Renal Function and Electrolyte Measurements

Reliable measurements of renal function markers and electrolytes are essential in many areas of patient care.

Duotrol® Normal can be used to monitor suitable methods for:

  • Creatinine
  • Urea
  • Uric acid
  • Sodium
  • Potassium
  • Chloride
  • Calcium
  • Magnesium
  • Phosphate

Regular quality control helps detect analytical changes that could affect the interpretation of laboratory results.

Glucose and Lipid Metabolism

Routine clinical chemistry testing also includes measurements associated with metabolic health and cardiovascular risk assessment.

Relevant analytes include glucose, total cholesterol, HDL cholesterol, LDL cholesterol and triglycerides.

Quality control at a defined concentration level helps laboratories monitor the performance of these assays over time, particularly when analytical conditions or reagent lots change.

Proteins and Additional Laboratory Parameters

Depending on the method and instrument, Duotrol® Normal also provides assigned values for albumin, total protein, immunoglobulins, C-reactive protein (CRP), ferritin and other clinical chemistry parameters.

Laboratories should always use the target values corresponding to their specific analytical methods.

Independent Quality Control Across Different Analyzer Systems

Duotrol® Normal is a manufacturer-independent quality control material for clinical chemistry testing.

Method- and instrument-specific assigned values are available for various analytical systems, including Beckman Coulter AU and Roche cobas analyzers, as well as selected measurements on Siemens Atellica systems.

This flexibility allows laboratories to integrate the control serum into established quality control procedures without relying exclusively on materials supplied by the analyzer manufacturer.

Independent control materials can provide an additional means of assessing assay performance, particularly when changes occur in reagents, calibrators or analytical conditions.

For further information on the role of independent controls in laboratory medicine, read our article on independent quality control, Rili-BÄK, IVDR and ISO 15189.

For methods without assigned values, laboratories must establish and verify suitable control ranges before routine use.

Lyophilized Human Plasma-Based Control Material

Duotrol® Normal is supplied as a lyophilized control serum based on defibrinated human plasma. The formulation also contains chemicals, preservatives and stabilizers.

Before use, each vial must be reconstituted with exactly 5 mL of distilled or deionized water.

After adding the water, close the vial and allow it to stand for 30 minutes, protected from light. Gently swirl the vial occasionally to support complete dissolution, taking care to avoid foaming.

Once properly reconstituted, the control material can be processed using the appropriate clinical chemistry measurement procedure.

Quality Control for Quantitative in vitro Diagnostic Measurements

Duotrol® Normal is intended for quality control of quantitative diagnostic measurements performed in vitro.

The control results are assessed against the assigned target values and the applicable control limits for each analyte and analytical method.

Lot-specific and method-specific values provide a basis for evaluating analytical performance. The stated control ranges reflect the maximum permissible deviations specified under the German Medical Association's laboratory quality assurance guideline (Rili-BÄK).

Laboratories should also establish their own statistical control limits according to their quality management procedures and applicable requirements.

The product name “Normal” identifies the control level. It does not imply that every assigned analyte concentration corresponds to a physiological reference interval.

Using Duotrol® Normal and Duotrol® Abnormal Together

Duotrol® Normal and Duotrol® Abnormal are complementary control materials covering different concentration levels.

Duotrol® Normal supports routine monitoring at the lower control level, while Duotrol® Abnormal provides a higher control level for many clinical chemistry analytes.

Using both levels allows laboratories to assess analytical performance across different concentration ranges rather than relying on a single control point.

This approach can help identify concentration-dependent analytical variation and supports a more comprehensive internal quality control strategy.

Storage Conditions and Stability After Reconstitution

Unopened Duotrol® Normal should be stored at +2 °C to +8 °C until the stated expiry date.

For most analytes, the following stability periods apply after reconstitution, provided that the material is protected from light:

  • +15 °C to +25 °C: 12 hours
  • +2 °C to +8 °C: 5 days
  • −15 °C to −25 °C: 28 days

Reconstituted material must not be frozen more than once.

Some analytes have shorter stability periods, particularly direct and total bilirubin, acid phosphatase and ALT. The applicable analyte-specific stability limits must be observed.

Duotrol® Normal – Product Specifications

Product type: Lyophilized quality control serum

Application: Internal quality control in clinical chemistry

Control level: Lower control level

Material: Defibrinated human plasma-based control

Measurement type: Quantitative clinical chemistry assays

Compatibility: Suitable clinical chemistry analyzers using applicable assigned values

Reconstitution: 5 mL distilled or deionized water

Pack size: 1 × 5 mL

Unopened storage: +2 °C to +8 °C

Reliable Routine Quality Control with Duotrol® Normal

Duotrol® Normal supports the ongoing assessment of analytical performance in medical laboratories. Its broad analyte coverage and method-specific assigned values make it a practical option for routine quality control across different clinical chemistry assays.

By monitoring results at a defined lower control level, laboratories can evaluate assay consistency and identify changes that may require investigation.

Combined with Duotrol® Abnormal, the control serum forms part of a two-level quality control approach for monitoring quantitative clinical chemistry measurements across different concentration ranges.

Frequently Asked Questions About Duotrol® Normal

What is Duotrol® Normal used for?

Duotrol® Normal is a lyophilized quality control serum used to monitor the precision and trueness of quantitative clinical chemistry measurements at the lower of two complementary control levels.

Which analytes can be monitored with Duotrol® Normal?

Depending on the method and instrument, Duotrol® Normal provides assigned values for enzymes, electrolytes, renal function markers, glucose, lipids, proteins and other clinical chemistry analytes.

Is Duotrol® Normal compatible with different clinical chemistry analyzers?

Yes. It is a manufacturer-independent control material with assigned values for selected methods on different analyzer systems. Laboratories must use the values appropriate to their instruments and analytical methods.

How do I reconstitute Duotrol® Normal?

Add exactly 5 mL of distilled or deionized water to the vial. Close it and allow it to stand for 30 minutes, protected from light. Swirl gently from time to time and avoid foaming.

What is the difference between Duotrol® Normal and Duotrol® Abnormal?

Duotrol® Normal represents the lower control level. Duotrol® Abnormal provides higher concentrations for many analytes, allowing laboratories to monitor analytical performance at an additional control level.

How long is Duotrol® Normal stable after reconstitution?

For most analytes, the reconstituted control is stable for five days at +2 °C to +8 °C or 28 days at −15 °C to −25 °C, with only one freezing cycle permitted. Some analytes have shorter stability periods.

More Information
Parameter Albumin, ALT / GPT, ALP / AP, Amylase, ASL-O, AST / GOT, Bilirubin, Bilirubin Direct, Bilirubin Total, Chlorid, Cholesterol, CHE, CK-NAC, CK-MB, Creatinine, CRP, Iron, Ferritin, Total Protein, gGT, Glucose, GLDH, Uric Acid, Urea, HBDH, HDL-Cholesterol, IgA, IgG, IgM, Potassium, Lactate, LDH, LDL-Cholesterin, Lipase, Lithium, Magnesium, Sodium, Osmolality, Pankreas-Amylase, Phosphat, Acid Phosphatase, Transferrin, Triglyceride
Packaging unit 1 x 5 mL Lyophilisat

If required, we can provide you with device-specific target values using statistical analysis. Information on participating in target value determination can be found here.

A product-specific IFU, including current target values, safety data sheets (MSDS), data sheets instead of MSDS, and CoAs are available as PDF downloads directly from our database: https://dokumente.biomed.de/

You will find your login details on your delivery note.

are usually dispatched on the same day (no dispatch on Fridays)
Orders placed before 12 noon
are usually dispatched on the same day (no dispatch on Fridays)
in an environmentally friendly box
Refrigerated shipping
in an environmentally friendly box
from 2.000 € Germany-wide
Free shipping
from 2.000 € Germany-wide