Rota-Adeno-Norovirus Test – GI/GII & Positive Control
Qualitative rapid test for the simultaneous detection and differentiation of rotavirus, adenovirus and norovirus genogroup I (GI) and II (GII) antigens in human stool specimens.
Rota-Adeno-Norovirus Rapid Test – Combined Antigen Detection with External Positive Control
The BIOMED Rota-Adeno-Norovirus Rapid Test is a qualitative immunochromatographic in vitro diagnostic assay for the simultaneous detection of rotavirus, adenovirus and norovirus genogroup I (GI) and genogroup II (GII) antigens in human stool specimens.
The assay enables the differentiated detection of four viral targets using a dual-strip test cassette. Separate detection regions for rotavirus, adenovirus, norovirus GI and norovirus GII provide visually interpretable results within 15 minutes, supporting the laboratory investigation of suspected viral gastroenteritis.
A key advantage of the BIOMED test kit is the included external Rota-Adeno-Norovirus positive control. This combined control contains material for all four detection targets, allowing additional verification of the assay's positive detection reactions.
Each kit contains 20 dual-strip test cassettes and the necessary components for sample preparation, making it suitable for professional use in clinical laboratories, hospitals and other diagnostic healthcare facilities.
What Are the Key Benefits of the Rota-Adeno-Norovirus Rapid Test?
- Four detection targets in one cassette: Rotavirus, adenovirus, norovirus GI and norovirus GII.
- Differentiated norovirus detection: Separate test lines for norovirus genogroups I and II.
- Results in 15 minutes: Rapid visual interpretation without a dedicated analyser.
- External positive control included: One combined control covering all four detection targets.
- Only one stool specimen required: A single sample preparation procedure for both test strips.
- 20 tests per kit: Includes sample collection tubes, dilution buffer and disposable pipettes.
- Supports gastroenteritis diagnostics: Simultaneous investigation of important viral pathogens associated with acute gastrointestinal infections.
- Storage at +2 to +30 °C: No refrigeration required when stored under the specified conditions.
Why Is Combined Rotavirus, Adenovirus and Norovirus Detection Clinically Relevant?
Acute viral gastroenteritis can be caused by different pathogens that frequently produce overlapping clinical symptoms. Common manifestations include sudden-onset diarrhoea, vomiting, nausea and abdominal discomfort.
Identifying the causative pathogen based on clinical presentation alone is therefore often difficult.
The combined Rota-Adeno-Norovirus antigen test enables several clinically relevant gastrointestinal viruses to be investigated in a single testing procedure. This can be particularly useful when evaluating acute diarrhoeal illness or suspected infectious gastroenteritis outbreaks.
Norovirus GI and GII – Separate Detection of Two Genogroups
Noroviruses belong to the family Caliciviridae and are among the major causes of acute viral gastroenteritis.
Due to their high infectivity, noroviruses can spread rapidly in environments involving close interpersonal contact, including hospitals, long-term care facilities, childcare centres and other communal settings.
Noroviruses associated with human gastroenteritis predominantly belong to genogroups GI and GII.
The BIOMED Rota-Adeno-Norovirus Rapid Test features separate detection regions for both genogroups. This enables differentiation between positive norovirus GI and norovirus GII antigen results.
Important: Differentiation is performed at genogroup level. The assay does not identify individual norovirus genotypes.
Rotavirus – Antigen Detection in Suspected Viral Gastroenteritis
Rotaviruses are important causes of acute gastrointestinal infections, particularly in infants and young children.
The assay detects the VP6 antigen of group A rotavirus. Qualitative antigen detection supports the diagnostic investigation of suspected rotavirus infection in patients presenting with compatible gastrointestinal symptoms.
Adenovirus – Additional Detection of Gastrointestinal Viral Pathogens
Certain adenovirus types, particularly enteric types 40 and 41, are associated with gastrointestinal infections.
The adenovirus detection region of the BIOMED assay targets the adenovirus hexon antigen.
Including adenovirus detection broadens the diagnostic scope compared with testing for rotavirus and norovirus alone.
External Rota-Adeno-Norovirus Positive Control Included
Verification of assay performance is an essential component of quality assurance in clinical laboratory diagnostics.
The BIOMED Rota-Adeno-Norovirus Rapid Test combines integrated procedural controls with an additional external positive control.
Integrated Procedural Controls
Each of the two test strips features its own green control line (C).
The appearance of these lines indicates that sufficient sample volume has been applied, the required lateral flow has occurred and the internal procedural control reactions have functioned correctly.
One External Positive Control for Four Detection Targets
The external positive control supplied with the kit consists of a control swab containing a combination of:
- Purified rotavirus viral proteins
- Purified adenovirus viral proteins
- Recombinant, non-infectious virus-like particles of norovirus genotype GI.1
- Recombinant, non-infectious virus-like particles of norovirus genotype GII.4
The positive control is prepared in the same manner as a patient specimen and applied to both test strips.
When the control is tested successfully, the rotavirus-adenovirus strip displays the red rotavirus test line, the blue adenovirus test line and the green control line.
The norovirus strip displays red test lines for GI and GII together with its green control line.
This allows the positive detection reactions of all four assay targets to be checked.
Your advantage: The combined external positive control is already included in the BIOMED test kit and does not need to be purchased separately.
How Does the Rota-Adeno-Norovirus Rapid Test Work?
The assay is based on a qualitative immunochromatographic lateral flow method.
The dual-strip cassette contains two separate reactive test strips.
Rotavirus and Adenovirus Test Strip
This strip uses specific monoclonal antibodies to detect rotavirus and adenovirus antigens.
Rotavirus detection targets the VP6 antigen of group A rotavirus, while adenovirus detection is based on recognition of the hexon antigen.
Norovirus GI and GII Test Strip
The second strip contains specific antibodies directed against norovirus antigens belonging to genogroups GI and GII.
Separate detection regions enable qualitative differentiation between these two genogroups.
Following sample preparation, the diluted stool specimen is applied to both test strips. Target antigens react with the specific antibodies, producing coloured test lines through immunochromatographic migration.
How Are the Test Results Interpreted?
Results are visually interpreted after 15 minutes.
The individual detection regions are identified as follows:
- Rotavirus positive: A red test line appears at R.
- Adenovirus positive: A blue test line appears at A.
- Norovirus GI positive: A red test line appears at GI.
- Norovirus GII positive: A red test line appears at GII.
- Valid test: A green control line (C) appears on each of the two test strips.
Several positive test lines may appear simultaneously, indicating the detection of multiple viral antigens within the same specimen.
A negative result is indicated when only the two green control lines are visible.
If a required control line is absent or irregular or non-specific colour changes occur, the result is invalid. Testing must be repeated using a new test cassette.
Test lines appearing after the 15-minute reading time have no diagnostic value.
What Do Positive Norovirus GI and GII Results Mean?
A positive norovirus GI result indicates the presence of detectable norovirus genogroup I antigens in the examined stool specimen.
A positive norovirus GII result indicates the detection of genogroup II antigens.
If both test lines are visible, antigens associated with both genogroups have been detected.
However, the assay does not differentiate individual genotypes within either genogroup. Due to the considerable genetic and antigenic diversity of noroviruses, not all circulating genotypes may be detected with equal sensitivity.
In patients with compatible clinical symptoms, a positive antigen result supports the suspicion of norovirus infection. However, the test does not provide information about viral load or patient infectivity.
If antigen testing is negative despite persistent clinical suspicion, additional molecular diagnostic testing may be appropriate.
Which Stool Specimens Can Be Tested?
The BIOMED Rota-Adeno-Norovirus Rapid Test is intended for the analysis of human stool specimens.
Suitable specimen consistencies include:
- Solid stool specimens
- Semi-liquid stool specimens
- Liquid stool specimens
Fresh, untreated specimens are recommended for optimal test performance.
Specimen collection should preferably take place as early as possible after the onset of diarrhoea and vomiting, as viral shedding may be particularly high during the early stages of infection.
If immediate testing is not possible, stool specimens may be stored at +2 to +8 °C for one to two days. Longer storage at −20 °C is possible, although changes in antigen immunoreactivity cannot be excluded.
Repeated freezing and thawing should be avoided.
Specimens containing added preservatives, transport media or enrichment media are unsuitable for testing.
Heavily blood-contaminated stool specimens may produce non-specific reactions and affect result reliability.
How Reliable Is the Rota-Adeno-Norovirus Rapid Test?
The diagnostic performance of the BIOMED Rota-Adeno-Norovirus Rapid Test has been evaluated for all four detection targets.
Diagnostic Sensitivity and Specificity
Rotavirus
- Sensitivity: Greater than 99.9%
- Specificity: 99%
Adenovirus
- Sensitivity: Greater than 99.9%
- Specificity: 98%
Norovirus GI
- Sensitivity: 88%
- Specificity: 99%
Norovirus GII
- Sensitivity: 95%
- Specificity: 97%
PCR was used as the reference method for rotavirus and norovirus, while adenovirus performance was evaluated against a commercially available rapid test.
The reported performance characteristics must therefore be interpreted in relation to the respective comparison methods.
Analytical Detection Limits
Analytical sensitivity was established at the following concentrations:
- Rotavirus: Approximately 31 ng/mL
- Adenovirus: Approximately 31 ng/mL
- Norovirus GI.1: 12.5 ng/mL
- Norovirus GII.4: 1.5 ng/mL
Recombinant virus-like particles of norovirus genotypes GI.1 and GII.4 were used to determine the norovirus detection limits.
These values describe performance under the evaluated conditions and cannot necessarily be extrapolated to all norovirus genotypes.
Repeatability and Reproducibility
Repeatability testing using positive, low-positive and negative control concentrations demonstrated 100% agreement within a single testing day.
Additional evaluations across multiple testing days and operators confirmed reproducibility of the qualitative detection method under the conditions investigated.
Interference and Cross-Reactivity
The assay was evaluated for potential interference from various substances and cross-reactivity with other microorganisms.
Under the investigated conditions, no relevant interference was observed at the tested concentrations of selected medications and specimen constituents.
Likewise, no corresponding cross-reactivity was observed with the gastrointestinal bacteria, viruses, fungi and parasites evaluated.
These findings apply to the specific substances, concentrations and microorganisms investigated.
What Are the Diagnostic Limitations of the Test?
The Rota-Adeno-Norovirus Rapid Test is a qualitative antigen assay and does not provide quantitative measurements of viral concentration.
A negative result does not reliably exclude rotavirus, adenovirus or norovirus infection.
Potential causes of false-negative results include:
- Antigen concentrations below the detection limit
- Inappropriate specimen collection or storage
- Specimen collection at an unsuitable stage of infection
- Norovirus genotypes that are not adequately recognised by the assay
Following recent oral vaccination with RotaTeq, positive rotavirus test results may occur within 15 days of vaccination.
A positive result also does not exclude additional infections caused by other gastrointestinal pathogens.
Final diagnostic interpretation must be performed by healthcare professionals in conjunction with clinical symptoms and, where appropriate, additional laboratory findings.
What Is Included in the Rota-Adeno-Norovirus Rapid Test Kit?
Each kit contains:
- 20 Rota-Adeno-Norovirus dual-strip test cassettes
- 20 sample collection tubes containing dilution buffer
- 20 non-graduated disposable pipettes
- 1 combined external Rota-Adeno-Norovirus positive control
The positive control contains material for rotavirus, adenovirus and norovirus GI and GII.
Additional materials required for testing include disposable gloves, a vortex mixer and a timer.
Storage and Stability
The Rota-Adeno-Norovirus Rapid Test can be stored in its original, unopened packaging at temperatures between +2 and +30 °C until the indicated expiry date.
Do not freeze the test components or external positive control.
Once the aluminium pouch has been opened, the test cassette must be used within 2 hours.
Frequently Asked Questions About the Rota-Adeno-Norovirus Rapid Test
Which Viruses Does the Rota-Adeno-Norovirus Rapid Test Detect?
The assay qualitatively detects rotavirus, adenovirus and norovirus genogroup I (GI) and II (GII) antigens in human stool specimens. Four separate detection regions enable differentiated interpretation.
Can the Test Differentiate Between Norovirus GI and GII?
Yes. The dual-strip cassette contains separate test lines for norovirus GI and GII. However, further genotyping within these genogroups is not possible.
How Quickly Are Results Available?
Results can be visually interpreted after 15 minutes. No dedicated analyser is required.
Is an External Positive Control Included in the Kit?
Yes. Each kit contains one combined external positive control covering rotavirus, adenovirus and norovirus GI and GII. It allows additional verification of all four positive detection reactions.
Can Multiple Viruses Be Detected Simultaneously?
Yes. The assay features separate detection regions. If the corresponding antigens are present, multiple test lines may become positive within the same testing procedure.
Is the Rota-Adeno-Norovirus Test Suitable for Investigating Gastroenteritis Outbreaks?
The assay can support laboratory investigations of suspected viral gastroenteritis. In particular, separate detection of norovirus GI and GII may provide additional diagnostic information.
Molecular diagnostic methods may be required for further outbreak investigation and epidemiological characterisation.
Does a Negative Result Rule Out Norovirus Infection?
No. A negative result does not completely exclude norovirus infection. Due to the antigenic diversity of noroviruses, certain genotypes may not be adequately detected. PCR testing may be appropriate when clinical suspicion persists.
Which Specimens Are Required?
The assay requires a human stool specimen. Solid, semi-liquid and liquid stool specimens can be analysed.
Who Is the Rota-Adeno-Norovirus Rapid Test Intended For?
The test is intended exclusively for professional in vitro diagnostic use by appropriately trained healthcare personnel.
Order the Rota-Adeno-Norovirus Rapid Test from BIOMED
The BIOMED Rota-Adeno-Norovirus Rapid Test combines qualitative detection of rotavirus, adenovirus and norovirus GI and GII in a single dual-strip cassette.
With four differentiated detection targets, 15-minute results and an included external positive control, the test kit provides a practical solution for professional viral gastroenteritis diagnostics.
For laboratories requiring only rotavirus and adenovirus antigen detection, BIOMED also offers the Rota-Adenovirus Rapid Test.
Explore our range of Point-of-Care Rapid Tests for additional diagnostic solutions in gastrointestinal and respiratory testing.
| Parameter | Adenovirus antigen, Norovirus genogroup I antigen, Norovirus genogroup II antigen, Rotavirus antigen |
|---|---|
| Packaging unit | 20 x test cassettes, 20 x dilution buffers and 1 x positive control |
Watch our video instructions to learn how to properly handle the Rota-Adeno-Norovirus Test:
1. Teil: Rota-Adeno-Norovirus Test positive control
2. Teil: Rota-Adeno-Norovirus Test
2. Teil: Rota-Adeno-Norovirus Test results and interpretation
Before using the rapid tests for the first time, we recommend that you watch all three videos in the correct order.